Chapter 3 81
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Amplitude
Amplitude specifications do not apply for the negative peak detector
mode.
Specifications Supplemental Information
Measurement Range Displayed Average Noise
Level to Maximum Safe Input
Level
Input Attenuator Range 0 to 60 dB, in 5 dB steps
Specifications Supplemental Information
Maximum Safe Input Level
Input attenuator setting
15 dB Signals > +33 dBm (2 W) or
+79 dBmV (1W) (Option 1DP)
nominal may trigger input
protection, which disconnects
the input path.
Average Continuous Power
or Peak Pulse Power
50 +30 dBm (1 W)
75 (Option 1DP) +75 dBmV (0.4 W)
dc 100 Vdc dc tr ansients may
momentarily trigger input
protection.
Input attenuator setting
<15 dB Signals > +6 dBm (4 mW) or
+61 dBmV (15mW)
(Option 1DP) nominal may
trigger input protection,
which automatically increases
input attenuation to 15 dB.
Average Continuous Power
or Peak Pulse Power
50 +3 dBm (2 mW)
75 (Option 1DP)+59 dBmV (10 mW)
dc 100 Vdc dc transients may trigger
input protection.