56 Chapter 2
Agilent E4408B Specifications and Characteristics
Amplitude
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Informati on
Spurious Responses
Second Harmonic Dis t o rti on
Input Signal
10 MHz to 500 MHz < −60 dBc f o r −30 dB m signal
at input mixera+30 dBm SHI (second
harmonic intercept)
500 MHz to 1.5 GHz < −70 dBc for −30 dBm signal
at input mixera+40 dBm SHI
1.5 GHz to 2.0 GHz < −80 dBc for −10 dBm signal
at input mixera+70 dBm SHI
2.0 GHz to 3.35 GHz < −95 dBcb for −10 dBm signal
at input mixera
+85 dBm SHI
3.35 GHz to 6.6 GHz < −95 dBcb for −10 dBm signal
at input mixera
+85 dBm SHI
6.6 GHz to 13.25 GHz < −95 dBcb for −10 dBm signal
at input mixera
+85 dBm SHI
Third Order Intermodulation
Distortion
10 MHz to 100 MHz +5 dBm TOI (third order
intercept), charact eristic
100 MHz to 3 GHz < −75 dBc for two −30 dBm
signals at input mixera and
>50 kHz separation
+7.5 dBm TOI
3.0 GHz to 6.7 GHz < −75 dBc for two −30 dBm
signals at input mixera and
>50 kHz separatio n
+7.5 dBm TOI
6.7 GHz to 13.2 GHz < −70 dBc for two −30 dBm
signals at input mixera and
>50 kHz separatio n
+5.0 dBm TOI
13.2 GHz to 26.5 GH z < −70 dBc for two −30 dBm
signals at input mixera and
>50 kHz separatio n
+5.0 dBm TOI
Other Input Related Spurious
Inband Responses
>30 kHz offset < −65 dBc for −20 dBm signal
at input mixera