Getting Started and Using Masks

To align a Mask to the waveform

To align a Mask to the waveform

1Access the Mask Test Setup dialog box and make sure the mask test is enabled.

2Locate the Use File Setup When Aligning option.

When aligning a mask to the waveform, these settings are used:

Channel range and offset

Time base range and position

Trigger mode and level

Mask test scaling source, position(s), and delay

You can either use the current oscilloscope settings, or you can use the standard settings defined in the mask file.

3If you want to use the standard settings defined in the mask file, checkmark the Use File Setup When Aligning option.

4If you want to use the current oscilloscope settings, do not checkmark this box.

When aligning a mask, you can use the current settings for the oscilloscope setup, or use the standard settings in the mask file.

5Click the Align button to align the mask.