System Configuration

The basic 8510C measurement system configura- tion needed for test port flatness correction is shown in Figure 1. A single channel power meter such as the Agilent 437B, E4418A, or E4418B is required, along with a compatible sensor, for per- forming the power flatness calibration. Power meters with the 437B command set must be used for the power level flatness correction.

 

85107B Measurement System

 

8510C network analyzer

 

8517B test set

bus

 

system

8485A

power sensor

 

8510

E4418B

power meter

 

 

83651B

 

synthesized sweeper

Figure 1. Basic Agilent 85107B measurement system con- figuration for test port flatness correction

The Agilent 8511B frequency converter test

set may be used with the test port power flatness- correction feature. The range of the test port power is determined by the available power of the 8360 source and the insertion loss of the com- ponents added to the measurement system.

3

Page 3
Image 3
Agilent Technologies PN 8510-16 manual 85107B Measurement System, System Configuration