SECTION 7 PERFORMANCE TEST
7-5

7.3.2 Output level frequency characteristics

(1) Test specifications
Frequency characteristics with reference to 0 dBm
±0.5 dB
(2) Measuring instrument for test
Power meter (100 kHz to 2080 MHz)
(3) Setup

Fig. 7-2. Output Level Frequency Characteristic Test

(4) Test procedure
STEP PROCEDURE
1. Adjust the power meter to the zero point and correct the sensor sensitivity.
2. Set the MG3641A/MG3642A output level to 0 dBm.
3. Set the MG3641A/MG3642A to the frequency to be measured.
4. Set the calibration factor of the sensor of the power meter and read the output level.
5. Repeat steps 3 and 4.
(5) Note on test
The MG3641A/MG3642A output level is defined with the RF output connector. To measure the output level,
directly connect the power sensor to the MG3641A/MG3642A RF output connector.
Sensor Power meter
SENSOR
INPUT
MG3641A/MG3642A