SECTION 8 CALIBRATION

8-1

SECTION 8CALIBRATION

8.1 Calibration Required

The calibration is performed as the preventive maintenance to prevent the MG3641A/MG3642A performance from
reducing.
The calibration is required when the inside of this device need periodically be adjusted to maintain the MG3641A/
MG3642A performance even if the MG3641A/MG3642A itself operates normally.
If an item that does not satisfy the calibration standard is detected, contact out service division.

8.2 Calibration Device List

Table 8-1 lists the calibration devices according to the calibration items.

Table 8-1. Calibration Device List

Calibration item Measuring Requested Recommended
instrument performance* device
Reference oscillator Oscilloscope 10 MHz measurable
frequency accuracy External trigger possible
Reference oscillator Frequency standard Standard radio wave receiver
frequency accuracy or equivalent (Accuracy:
1 × 10-9 order or more)
* Extracts a part of performance that can cover the calibration item measurement range.