ChannelMag Section 2 Installation
2.3.Straight Lengths of Channel / Pipe
2.3.1Straight Lengths of Channel (CM2)
2.3.2Straight Lengths of Pipe (PM2)
the electrode cables and the energizer coil cable within the junction box. Do not use silicone rubber (RTV) for potting, since it contains acetic acid which can corrode the junction box terminals.
Make sure the serial number of the ChannelMag sensor(s) shown on the cables agrees with the serial number of the transmitter. This is particularly important when more than one ChannelMag flow system has been supplied. If a sensor and transmitter have been supplied separately, then the transmitter range factor has to be adjusted according to the sensor calibration (see transmitter instructions).
Although the ChannelMag sensor is
The CM2 ChannelMag must be installed in a channel with a minimum of 7 straight channel widths upstream and 4 straight channels downstream, as shown in the diagram below. For
The straight lengths are measured from the end of the ramps. When ramps are not supplied (and the ChannelMag CM2 sensor(s) is inserted into a recess in the base of the channel), then the straight lengths are measured from the end of the CM2 sensor.
W = channel width (not necessarily the width of the CM2 sensor)7W = upstream minimum length2W = downstream minimum length (must be 7W for
The ChannelMag and installation conforms to German DIN 19559 and German Waste Water Institute (ATV Arbeitsgruppe 1.2.5) requirements.
The ideal location in the pipe is where there is a maximum straight length. The table below provides recommendations of minimum straight lengths of pipe for various pipe configurations. Shorter lengths or other configurations affect published accuracy, dependant on pipe size and velocity range. Consult ARKON or a qualified representative.
Diametrically opposite sensors are either 1 pair or 2 pair.
When an ultrasonic level transducer is used for partially filled pipe applications it must be located at least 8” (200 mm) before or after the ends of the ramps to avoid drops in level due to
10