Display settings
88 Nio® medical display systems
• To adjust display luminance manually, adjust the Luminance
slider. This function is not available for MFCD/MFGD1218
displays.
The luminance setting is saved when the system is shut
down.
If the slider is positioned above the CAL mark, the luminance
is in calibrated position, as determined during the latest
calibration. If the slider is not above the CAL mark, the
luminance is not in calibrated position.
Important
If you want the system to be DICOM compliant, you must select
the DICOM display function and calibrate the display. After
calibration, the luminance must remain in calibrated position to
maintain DICOM compliance.