Display settings
Nio® medical display systems 93
Figure 47: DICOM Optimizer dialog
2. Adjust the slider until there is almost no visible difference
between the background of the image above the slider and
the bitmap inside this image.
3. When done, click button Set point 1 of 20.
4. Repeat this procedure until the button is grayed. This
indicates the last point is set.
5. Click OK to finish.
To reset to the latest calibration:
1. In the Calibration dialog (see step no. 4. in the description of
the calibration with sensor), check the option Use the latest
calibration.
2. Click the OK button.
The calibration and the luminance will be reset to the latest
calibrated position.
Note:
When you start up the system, the calibration is also reset to
the latest calibrated position. The luminance, however, is
kept at the latest value.