What You See | What It Means | What You Do |
| |
E11 or higher | Meter Malfunction | Turn meter off, then on | ||
|
|
| using . Retest. If error | |
|
|
| remains, contact the | |
|
|
| Bayer Diabetes Helpline | |
|
|
| (see pg. 45). |
|
|
|
|
| |
The display goes | 1. You may have | 1. Press the | button. | |
blank when the | accidentally pressed |
|
| |
meter is turned | the | button. |
|
|
on. | 2. Meter will turn off after | 2. Press the | button. | |
| 3 minutes without use or |
|
| |
| 15 minutes of no use |
|
| |
| while waiting for you to |
|
| |
| apply blood or control |
|
| |
| solution. This is a normal |
|
| |
| expected action that |
|
| |
| saves battery life. |
|
| |
|
|
| ||
Meter will not | 1. Battery life has ended. | 1. Replace Battery | ||
turn on. |
|
| (see pg. 28). |
|
| 2. Battery holder is loose or | 2. Check the battery and | ||
| battery is installed + side | the battery holder to be | ||
| down. |
| sure they are tightly in | |
|
|
| place and correctly | |
|
|
| installed (see pg. 28). If | |
|
|
| display is still blank, | |
|
|
| replace battery (CR2025). | |
|
|
| ||
Meter fails to | 1. Not enough blood. | 1. Discard used test strip. | ||
start countdown |
|
| Retest using a round | |
after blood is |
|
| drop of blood. | |
applied. | 2. The inside of your meter | 2. Clean inside of meter | ||
| may be dirty, or foil | (see pg. 27). Retest. | ||
| from the |
|
| |
| may be lodged in the |
|
| |
| front of the meter. |
|
| |
|
|
|
|
|