Marking a Control Test

Marking a test result as a control test excludes it from the 14-day average.

You must mark the result immediately after the test is performed—while the result is still displayed on the screen. To do this, follow these steps:

5. Open the Button Door.

Press the button to mark as a control.

The “Control” marker looks like this:

Press to accept.

6.To release the used test strip, hold the meter with the test strip pointing down

over a waste container. Press the release button. Make sure the heel of your hand is not against the meter handle. Note: If you block the meter handle with your hand, the test strip may not release properly.

7. Press the

to turn your meter

 

off.

15