Normal Control Solution

 

 

 

 

Total

Total

Strip

 

No. of

Mean

SD

CV

Lot

Operator

Assays

(mmol/L)

(mmol/L)

(%)

A

Lay User (n=50)

100

8.394

0.400

4.8

B

Lay User (n=50)

100

8.317

0.356

4.3

 

 

 

 

 

 

High Control Solution

 

 

 

 

Total

Total

Strip

 

No. of

Mean

SD

CV

Lot

Operator

Assays

(mmol/L)

(mmol/L)

(%)

A

Lay User (n=50)

100

18.706

0.961

5.1

B

Lay User (n=50)

100

18.867

0.800

4.2

 

 

 

 

 

 

Accuracy1.Bayer Accuracy Evaluation:

To assess the accuracy of the Ascensia® BREEZE ™ System, 102 fresh fingerstick specimens from people with diabetes were tested with the Ascensia® BREEZE ™ (2 readings per fingerstick) and the YSI 2300 STAT Plus Glucose Analyzer. The glucose concentration of the samples ranged from 2.3 to 30.1 mmol/L with an average of 9.9 mmol/L. The range of hematocrits was 29% to 55% with an average of 40%.

33