Error Codes and Symbols:
If you do not resolve the problem, contact Bayer Customer Service at 0845 600 6030 in the UK or 1 890 920 111 in Ireland.
| WHAT YOU SEE | WHAT IT MEANS | WHAT YOU SHOULD DO | |||||
|
|
|
|
|
|
|
|
|
|
|
| (will remain | Low batteries | Replace the batteries as | |||
|
|
| on the display) |
| soon as possible. | |||
|
|
|
|
|
|
|
|
|
|
|
| (will flash for 10 | Dead batteries | Replace the batteries. | |||
|
|
| seconds then |
|
| |||
|
|
| the meter will |
|
| |||
|
|
| turn off) |
|
| |||
|
|
|
|
|
|
|
|
|
| E1 |
|
|
|
| Temperature out | Move to an area that is | |
|
|
|
|
|
|
| of range | within the operating range |
|
|
|
|
|
|
|
| for the meter: 5°C – 45°C. |
|
|
|
|
|
|
|
| Let the meter adjust to this |
|
|
|
|
|
|
|
| temperature for twenty |
|
|
|
|
|
|
|
| minutes before testing. |
|
|
|
|
|
|
|
|
|
| E2 |
|
|
|
| The test strip has | Remove the test strip | |
|
|
|
|
|
|
| not filled enough | and retest using a new |
|
|
|
|
|
|
| for an accurate test. | test strip. |
|
|
|
|
|
|
|
|
|
| E3 |
|
|
|
| The meter is sensing | Remove the test strip and | |
|
|
|
|
|
|
| a used test strip. | retest using a new test |
|
|
|
|
|
|
|
| strip. Wait until you see the |
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| flashing blood drop on the |
|
|
|
|
|
|
|
| display before adding the |
|
|
|
|
|
|
|
| blood sample. |
|
|
|
|
|
|
|
|
|
| E4 |
|
|
|
| Test strip not | Remove and reinsert the | |
|
|
|
|
|
|
| inserted correctly. | test strip properly (see |
|
|
|
|
|
|
|
| page 8). |
|
|
|
|
|
|
|
|
|
| E5 |
| E12 | Potential software | Remove the test strip | |||
| E6 |
| E13 | or hardware issue | and retest using a new | |||
| E8 |
| E14 |
| test strip. If you continue | |||
| E9 |
|
|
|
|
| to have problems, contact | |
|
|
|
|
|
|
|
| Bayer Customer Service. |
|
|
|
|
|
|
|
|
|