Maintenance, Lubrication & Adjustments

Interior Adjustments

Photo Q

Overall View

Photo R

Full Out/Deployed

Full In/Stowed

Positional

Positional

Sensor

Sensor

Pick-Up

Pick-Up

Plate

Plate

Photo S

Full In/Full Out Position of Carriage: The carriage houses two sensors that provide positional feedback of full out (see Photo R) or full in/stowed (see Photo

S)for the carriage and platform assembly. The prox- imity sensors switch on and off as the sensor head passes over a metallic pick-up plate mounted on the cassette frame. Distance from the sensor head to the pick-up plate must be approximately 4 mm and should be checked (due to possible loosening) to

ensure sensor switching (sensor LED lights when switched on). Note: Verify the full out sensor is distanced properly to remain lit (switched on) during the entire elevation cycle. If the sensor light should be intermittent during the elevation cycle, reposition the sensor to reduce the distance between the sen- sor head and pick-up plate, while confirming there is no interference of the sensor head and pick-up plate during the in/out movement. See Photo R.