Bunn LCA-1, LCR-1 r un, Calibrating The Dispenser, Calibration Mode,  Description, En g, r u n

Models: LCA-1 LCR-1

1 23
Download 23 pages 51.01 Kb
Page 17
Image 17
CALIBRATING THE DISPENSER

CALIBRATING THE DISPENSER

NOTE: The LCA/LCR-1 is calibrated at the factory and does not normally need to be re-calibrated. Open the dispenser door to access the digital Programming Switches with LED display. To enter the Calibration Mode, Set the RUN/NIGHT/PROGRAM Switch to the PROGRAM position and then hold both Up and Down Arrow switches until “1 r u n” appears of the display, (approx 15 sec).

LED DISPLAY

11 r unr u n r En guCALIBRATION MODEn

1

2 2

2

r u n

2

7 6 0

3

4 0

4

En g

5

E _ _

CALIBRATION MODE

 DESCRIPTION

Pump Cal. Procedure

Set PRIME-NORMAL-RINSE switch

 

to PRIME position. Place Graduated

 

Cylinder under dispense tip. Press and

 

hold Dispense Switch until pump stops

 

(20 sec) Refer to Calibration Procedure.

Pump Cal. #

Enter mL collected from the 20 sec.

 

test above.

 

Default = 22

 

Range = 18 - 26mL

Water Flow Rate

Set PRIME-NORMAL-RINSE switch

 

to RINSE position. Place Graduated

 

Cylinder under dispense tip. Press and

 

hold Dispense Switch until pump stops

 

(20 sec) Refer to Calibration Procedure.

Water Flow Rate

Enter mL collected from the 20 sec.

 

test above.

 

Default = 760

 

Range = 780 - 820 mLiters

Cabinet Temperature

LCR-1 Only Used to set the Target

 

Cabinet Temperature

 

Default = 40°F (4°C)

 

Range = 35°F - 75°F (2° - 24°C)

English or Metric Mode

Used to select the desired units.

 

Default = English

 

S.I. for Metric

Error Code Memory

Display cycles through the last (3)

 

Error Codes.

 

Erase by holding the Decrease (-)

 

switches for 10 sec.

 

Currently have to hold (+) & (-)

17

44806.1 081211

 

Page 17
Image 17
Bunn LCA-1, LCR-1 service manual r un, Calibrating The Dispenser, Calibration Mode,  Description, En g, r u n, Led Display