Bunn LCR-3 service manual Possible Cause, Troubleshooting Procedures, No Display

Models: LCR-3

1 27
Download 27 pages 23.12 Kb
Page 22
Image 22

Troubleshooting

A troubleshooting guide is provided to suggest probable causes and remedies for the most likely problems encountered. If the problem remains after exhausting the troubleshooting steps, contact the Bunn-O-Matic Technical Service Department.

Only qualified service personnel should perform inspection, testing, and repair of electrical equipment.

Shorting the terminals or the application of external voltages to electronic components may result in component or circuit board failure.

Intermittent operation of electronic circuit boards is unlikely. board failure will normally be permanent. If an intermittent condition is encountered, the cause will likely be a switch contact or a loose wire connection at a terminal or crimp.

Solenoid removal requires interrupting the water supply to the valve. Damage may result if solenoids are energized for more than ten minutes without a supply of water.

The use of two wrenches is recommended whenever plumbing fittings are tightened or loosened. This will help avoid twists and kinks in the tubing.

Make certain that all plumbing connections are sealed and all electrical connections are tight and isolated.

This dispenser is heated at all times. Keep away from combustibles.

WARNING:

Exercise extreme caution when servicing electrical equipment.

Disconnect the brewer from the power source when servicing, except when electrical tests are specified.

Follow recommended service procedures.

Replace all protective shields or safety notices.

Screen Displayed

Possible Cause

Troubleshooting Procedures

 

 

 

NO DISPLAY

No Power to the Dispenser

Check Master ON/OFF Switch

 

 

Check Power Cord & Circuit Breaker

 

 

Service Required

 

 

 

 

 

 

HEATING

Water temp is below the preset

Wait for water temp to rise

Please Wait!

READY temp

Check Ready Temp. threshhold setting

 

 

Service Required

 

 

 

 

 

 

RINSE REQUIRED

Rinse Alarm Time Interval has

Rinse Dispenser, see Rinse Procedure

 

expired

Replace BIB, see Loading the Concentrate

 

 

 

 

 

 

XXXX BIB EMPTY

Concentrate BIB is empty.

Replace BIB, see Loading the Concentrate

Replace BIB Soon

BIB not properly connected.

Check BIB Connector Fittings

 

Tubing kinked or blocked.

Check Tubing Installation

 

Empty BIB Threshhold too high.

Check Empty BIB Threshhold, see Field

 

 

Calibrating the Epmty BIB Warning

 

 

 

 

 

 

____ TARGET RPM

The calculated Target RPM for

Check Pump & Flow Rate Calibration, refer

TOO HIGH !!

the pump indicated is out of

to Field Calibration Procedures

 

range

 

 

 

 

22

41129 052208

 

Page 22
Image 22
Bunn LCR-3 service manual Possible Cause, Troubleshooting Procedures, No Display, Calibrating the Epmty BIB Warning