2.Edit Exposure Mode Settings

2.3Setting DWT DEP Tab

2.3.1Setting DWT DEP for High-Density Field

1Set the DWT DEP for the High-Density Field effective

Touch [HIGH DENS.] to display the parameters. The appearance of the button switches over between the highlight (in white outline characters against a colored background) and normal display each time you touch the button. The parameters are available when the button is highlighted.

DWT DEP Screen with Parameters of HIGH DENS.

2

3

Set the STARTING DENSITY parameter of the High-Density Field

Touch the or button, and set the value within the range between 0 to 30.

If the parameters of the Low-Density Field are effective, this value cannot be set lower than the value of STARTING DENSITY of the Low-Density Field (P.12).

The value on the left is the setting saved previously, and the value on the right is the current setting.

The value indicates the lower limit of density to start compression as shown in the following table:

Setting Values for Starting Density of High-Density Field

Setting value

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

 

Starting density (D)

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

1.1

1.2

1.3

1.4

1.5

1.6

 

Setting value

15 16 17 18 19 20

21

22

23

24

25

26

27

28

29

30

Starting density (D)

1.7

1.8

1.9

2.0

2.1

2.2

2.3

2.4

2.5

2.6

2.7

2.8

2.9

3.0

3.1

3.2

Set the DEP WEIGHT parameter of the High-Density Field

Touch the or button, and set the value within the range between 0 to 20.

The value on the left is the setting saved previously, and the value on the right is the current setting.

The value indicates the compression ratio in the density area that is larger than the value of STARTING DENSITY.

Select a larger value for higher degree of compression, and select a smaller value for lower degree of compression.

– 10 –