3.QA Mode Settings

3

4

Display the MORE window

Touch [MORE] to open the window.

MORE Window

Set the LOW WEIGHT parameter

Touch the or button, and set the value within the range between 0 to 20.

The value on the left is the setting saved previously, and the value on the right is the current setting.

For details of the parameter, refer to the step 3 in 2.3.2 Setting DWT DEP for Low- Density Field (P.12) .

5

Set the LOW EFFECT parameter

Touch the or

button, and set the value within the range between 0 to 10.

 

The value on the left is the setting saved previously, and the value on the right is the cur-

 

rent setting.

 

 

For details of the parameter, refer to the step 4 in 2.3.2 Setting DWT DEP for Low-

 

Density Field (

P.12) .

NOTE: If the LOW EFFECT parameter is set to an excessive value, artifacts may occur, and the effective data of the exposure image may be damaged. Check the exposure image, and if the results of compression appear to be too strong, either set the parameter to a smaller value, or set the DWT DEP not effective.

6

Set the settings valid, and close the MORE window

Touch [OK].

– 25 –