3.QA Mode Settings

3.3Changing DWT DEP Tab

3.3.1Changing DWT DEP for High-Density Field

1Set the DWT DEP for the High-Density Field effective

Touch [HIGH DENS.] to display the parameters. The appearance of the button switches over between the highlight (in white outline characters against a colored background) and normal display each time you touch the button. The parameters are available when the button is displayed in highlight.

DWT DEP Screen with Parameters of HIGH DENS.

Representation of DWT DEP on a graph

A blue curve is always displayed on a graph, and when HIGH-DENS.or LOW- DENS. is effective, a white curve is added.

The blue curve represents the tone curve selected on the LUT tab, and the white curve represents the tone curve combining the blue and the DWT DEP tone curves.

In other words, when either of HIGH-DENS.or LOW-DENS.is effective, the original tone curve compensation and the DWT DEP tone curve compensation are performed at the same time.

2

Set the STARTING DENSITY parameter of the High-Density Field

Touch the or

button, and set the value within the range between 0 to 30.

 

The value on the left is the setting saved previously, and the value on the right is the cur-

 

rent setting.

 

 

For details of the parameter, refer to the step 2 in 2.3.1 Setting DWT DEP for High-

 

Density Field (

P.10) .

– 22 –