This manual should be used for initial set up of the system and for reference purposes.
1.1 GENERAL INFORMATION
The system is a true low-air-loss mattress system suitable for prevention and treatment of
Pressure Ulcers up through Stage IV.
The system has been tested and successfully approved with the following standards:
EN 60601-1
EN 60601-1-2
EN 55011 Class B
IEC61000-3-2
IEC 61000-3-3
FOR US AND CANADA ONLY
Medical Equipment-Air Blower
E228589with respect to electrical shock, fire and
53DGmechanical hazards only in accordance with
UL60601-1 AND CAN/CSA C22.2 NO.601.1
Le produit à ététesté avec des équipementsmédicauxetrespecte les normes UL 60601-1 &
CAN/CSA C22.2 No.601.1. prévenant les choc électrique, le feu et les risques de blessures
physiques.
1.2 INTENDED USE
This product is intended to help treat and reduce the incidence of pressure ulcers from Stage I –
Stage IV, while optimizing patient comfort. It is appropriate for use in home care, long term care
and acute care settings.
NOTE:Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
NOTE:L’équipement ne peutêtreutilisés’il y a risque de mélange d’un anesthésique
inflammable avec l’airoul’oxygèneouoxydenitreux.
3

2. PRODUCT DESCRIPTION

11
7.3 AIR FILTER REPLACEMENT
1. Replace the air filter locatedat the bottom of blower.
2.The filter is reusable and can be washed gentlywith a mild
detergent and water. Dry the filter before use.
3. Check and replace air filterregularly if environment is dirty.
7.4 RECHARGEABLE BATTERY
1. To check if the rechargeable battery has been drained out, unplug the power cord andsee if the
Power Failure indicator will light up along with buzzer last for a few minutes.
2. If the Power Failure Alarm is unable to work or the battery might need to be
replaced(approximate life expectancy 6 months),user can contact your dealer or notify the
technician for replacement.
8. TROUBLESHOOTING
Problem Solution
Power is not ON Check if the plug is connected to mains.
Check if there is any blown fuse.
Alarm is on
(Audible & visual)
Check if the Quick Connector is rotated to the Lock/on position.
Check if the power is suddenly shut down.
Check if the connection between air tube connector to blower unit is
tightly secured.
Check if all tubing connections along mattress are secured.
Patient is bottoming out
Pressure setting might be inadequate for thepatient, adjust c omfort
range 1 to 2 levels higherand wait for a few minutes for best
comfort.
Automatic fowler boost is
not functioning
Check if the fowler boost adaptor is connected and secured to the
blower unit.
Check if the fowler boost sensor pad tightly snapped to the
designated location inside the bottom of the mattress foam pocket.
If the above information does not solve your problems, please contact your local agent directly.They
might require a technician totake care the problem.