By wiping the mattress unit with a damp cloth pre-soaked with warm water (not to exceed 65 )
containing a mild detergent, and keep it away from dust. The cover may also be cleaned using
sodium hypochlorite diluted in water. All parts should be air dried thoroughly before use.
CAUTION: Do not use phenolic-based product for cleaning.
CAUTION: After cleaning, dry the mattress without direct exposure of sunlight.
The carrying bag (if available) should be turned inside out and completely wiped down using the
disinfectant solutions. Allow it to air-dry thoroughly. Once the inside is dry, turn it back and wipe
down the outside of the bag with disinfectant solutions.
6. STORAGE
1. To quickly vacuum air out from mattress for storage, rotate the Quick Connector to the CPR
position and press the “Max Firm” button. It will vacuum air out within 5 minutes and then blower
will stop automatically.
2. Disconnect the automatic fowler boost adaptor on the side of the blower unit.
3. Rotate Quick Connector to the “Disconnect” position and pull up the Quick Connector.
4. Lay the mattress out flat and upsides down.
5. Roll from the head end towards the foot end.
6. Foot-end strap can then be stretched around the rolled mattress to prevent unrolling.
7. The power cord could be wrapped around the blower bumper on the back of blower.
CAUTION: When the blower is vacuuming air, air temperature is hot. Do not block the air
outlet.
NOTE: Do not fold, crease or stack the mattresses.
7. MAINTENANCE
7.1 GENERAL
1. Check main power cord and plug, and auto fowler boost cord if there are abrasions or excessive
wears.
2.Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are
stubbed together correctly.
3. Check the air hoses for any kink or break. For replacement, please contact your local dealers.
7.2 FUSE REPLACEMENT
1.Disconnect the plug form mains power when a blown fuse is
suspected.
2. Remove the cover of the fuse holder by means ofa small
screwdriver.
3.Insert a new fuse of the correct rating in, and replace the
cover of the fuse holder back. The fuse should be rated as
5A/250V(for 120V system) and T3A /250V(for 230V system)
type.
2.1 BLOWER AND MATTRESS SYSTEM
1. Quick Connector
2. BlowerUnit
3.Differential Zone Mattress
4.Auto Fowler Boost Sensor Pad (for standard 35.4” mattress only)
5. Mattress Foam Pocket