Front
1. Quick Connector
2. Front Panel
3. Handle
4. Auto Fowler Boost Adaptor
Rear
1. Power Switch
2. Power Socket
3. Fuses
4. Mounting Bracket
5. Blower Bumper
6. Air Filter
2.3 FRONT PANEL
1. ComfortLevel
Comfortlevel controls the air pressure output. When turning up ( ), the output pressure will
increase and higherpressure output will support the heavier weight user. Vice versa( )for
decreasing air pressure.
NOTE:Maximum Weight Capacity : 200kg or 440 lbs (for 35.4”model)
450kg or 1,000lbs (for 48” model)
5
9
4.3 PRESSURE SET UP
Users can adjust pressure of air mattress to a desired softness by adjusting the Comfort Level.
Please consult your physician by the below weight support range for a suitable setting.
NOTE:Maximum Weight Capacity : 200kg or 440 lbs (for 35.4”model)
450kg or 1,000lbs (for 48” model)
4.4 AUDIBLE AND VISIBLE ALARM
When the following situation occurred, the LED lights will light up along with the buzzer.
4.4.1 CPR – When CPR or quick storage is performed, or the Quick Connector is on Disconnect
position under power on situation, the “CPR” indicator (orange LED) will light up
along with buzzer.
4.4.2 Power Failure – When electrical shortage occurred or power cord is disconnected without
turning off the blower first, the “power failure” indicator (orange LED) will
light up along with buzzer. Check to ensure power cord is connected
properly.
4.4.3 Low pressure – When an abnormal low pressure occurred in body section, the “Low
Pressure” indicator (orange LED) will light up along with buzzer. Check all
tubingsare correctly installed as per installation instructions and ensure
all tubingsare secured properly. The Low Pressure indicator will
continuously light up until the low-pressure fault condition is resolved.
4.5 ALARM MUTE
When these situations mentioned above occurred, both the LED light and buzzer will sound off to
warn the patient. By pressing the button, it will temporary mute the buzzer (CPR & Low Pressure)
so the caregiver may check for possible causes. Should the CPR &low pressuresituation not
resolved within 3 minutes, the alarm will resume to beep. When in Power Failure situation,
pressing alarm mute button, it will cease the buzzer and turn off the “power failure” indicator.
5. CLEANING
By wiping the blower unit with a damp cloth pre-soaked with a mild detergent, and keep it away
from dust. If other detergent is used, choose one that will have no chemical effects on the surface
of the plastic case of the blower unit.
CAUTION: The Fowler Boost Sensor Pad should be always kept away from water.
CAUTION: Do not immerse or soak blower unit.
10