Grounding:
Before any connection to the output connectors is made, the unit shall be connected to a protective
earth conductor via the three-core main cable; the mains plug shall be inserted only into a socket
outlet provided with a protective earth contact. The protective action shall not be negated by the
user of an extension cord without protective conductor.
Several installation tips are listed below:
After installation, the extra length of the power cord, if any, should be neatly arranged to avoid any
tripping accidents. The EQUIPMENT should be firmly placed at position where users/doctors can
access easily.
4. OPERATION
NOTE: Always read the operating instruction before use.
4.1 GENERAL OPERATION
1. Switch on the main power switch found from the side of
the blower and press on the control panel to turn on
the power.
2. The syst em will automatic ally go into “Max Fi rm” mode
for a few minutesof inflation.
NOTE: The mattress is equipped with differential
zone feature, therefore center section of
the mattress will be inflated quicker than
the head and foot sections.
3. The low-pressure indicator (orange LED) will light up
when the mattress is not fully inflated.
4. According to the weight and height of the patient, adjust
the pressure setting to the most suitable level without
bottoming out.
4.2 CPR
When CPR needs to be performed, quickly rotate the Quick
Connector to the CPR position to release air.
2. Panel Lock
Should the panel remain untouched for5minutes, the panel lock feature will lock the panel with
green LED light on to prevent accident from changing the setting during normal operation. To
unlock, simply by pressing the Panel Lock button for 3 seconds.
3.Max Firm
This mode will be forced to execute every time the power is turned on. This insures the blower
is able to reach its maximum operating pressure. Once the max pressure level is reached, the
blower will automatically switch into the previous selected comfort level. User can also use this
function as full mattress inflation during patient ingress/egress for better support. There is a 20
minutes auto-recovery mode to return back the previous selected comfort level.
4. On/off
Press On to turn on the unit. Press Off to turn off/standby the unit.
NOTE: The power switch on the side of blower must be turned on.
5. Fowler Boost
The automatic Fowler Boost featuresadditional supports to the patient during upright position to
prevent bottoming out. This function is automatically activated wh en the bed is raised and
deactivated when the bed is flat.
NOTE: This function is only available for 35.4” replacement mattress system.
NOTE: When the patient in the raised position, it is recommended that caregiver readjust
the patient sitting posture to prevent possible bottoming out.
6. Alarm Mute
Press alarm mute button to temporary suspend the CPR and low-pressure alarm. Should the
situation not resolved within 3minutes, the alarm shall resume tonotify the patient.
A. CPR
During the CPR situation, the caregiver can rotate the quick connector to perform CPR.
Once the quick connector is switched to CPR mode, CPR alarm will sound with LED light on
to notify the caregiver.
B.PFA (Power Failure Alarm)
During power failure situation, the Power Failure LED light will light on with buzzer. By
pressing the mute button to disable both buzzer and LED.
C.Low Pressure Indicator
When low pressure LED continuously lights up with buzzer, the pressure inside of air
mattress body section is abnormal, please check all tubings arecorrectly installed as per
installation instructions and ensure all tubings are secured prope rl y.
NOTE: The low pressure indicator will not be detected if air cells of mattress are purposely
disconnected.
6