2640A 4-Wire Resistance Measurement Specifications

A-15.

Tables A-15 to A-17 provide 2640A specifications for the 4-wire resistance measurement function. The 4-wire measurements use 2 input channels a decade apart, e.g., channels 4 and 14.

Table A-15. 2640A 4-Wire Resistance Temperature Coefficient

Specification

Temperature Coefficient

Accuracy at -20°C

Characteristic

Add 1/10th the 90 day specification per ºC above 28ºC or below 18ºC.

Multiply the -10°C to + 60°C accuracy specification by 2. After 1 hour warmup. For accuracy between -10°C and -20°C, interpolate linearly.

Table A-16. 2640A 4-Wire Resistance Range and Resolution Specifications

Range

300Ω

3kΩ

30kΩ

300kΩ

3MΩ

Resolution

Slow

Fast

 

 

1 mΩ

3 mΩ

10 mΩ

30 mΩ

100 mΩ

300 mΩ

1Ω

3Ω

10Ω

30Ω

Current Applied

1mA

100µA

10µA

10µA

1µA

Full Scale

Voltage

300mV

300mV

300mV

3.0V

3.0V

Maximum voltage

applied by instrument

3.5V

3.5V

3.5V

3.5V

3.5V

Range

Table A-17. 2640A 4-Wire Resistance Accuracy Specifications

 

 

Accuracy, 3σ + (% input + V)

 

 

 

 

 

 

 

 

18ºC to 28ºC

 

-10ºC to 60ºC

 

 

 

 

 

 

90 Day

 

1 Year

 

1 Year

 

 

 

 

 

 

Slow

Fast

Slow

Fast

Slow

 

 

Fast

300Ω

3kΩ

30 kΩ

300 kΩ

3 MΩ*

.015%+20 mΩ

.02%+.3Ω

.03%+3Ω

.1%+40Ω

.25%+800Ω*

.02%+80 mΩ

.02%+.8Ω

.04%+10Ω

. 2%+100Ω

.5%+10kΩ*

.02%+50 mΩ

.02%+.5Ω

.03%+5Ω

.1%+60Ω

.25%+1 kΩ*

.02%+120 mΩ

.02%+1.2Ω

.04%+15Ω

.2%+150Ω

.5%+1.5kΩ*

.084%+126 mΩ

.084%+1.26Ω

.126%+12.6Ω

.42%+168Ω

1.05%+3.36Ω

.084%+336 mΩ

.084%+3.36Ω

.168%+42Ω

.84%+420Ω

2.1%+4.2 kΩ*

*The 3 MΩ range is susceptible to the absorption of humidity under extreme conditions. If the instrument is operated normally within its specified temperature-humidity range, the 3 MΩ range meets its accuracy specifications. However, if the instrument is “soaked” at 50 °C, 90% RH the 3MΩ range may require 1 hour of “dry-out” time at 25 °C, <40% RH for each hour of “soak” time in order to achieve its specified accuracy.

A-12

Page 222
Image 222
Fluke 2645A 2640A 4-Wire Resistance Measurement Specifications, Resolution Slow Fast, Current Applied, Full Scale Voltage

2645A, 2640A specifications

The Fluke 2640A and 2645A are advanced bench multimeters that represent a significant leap in measurement technology, designed primarily for electrical engineers and technicians who require precision and reliability in their readings. These multimeters are built to cater to complex applications in manufacturing, research, and development.

One of the standout features of the Fluke 2640A and 2645A is their high accuracy. With specifications that can reach up to 0.0035% DC Voltage accuracy, users can trust that their measurements will be consistent and precise. This is especially critical in fields like automated test equipment (ATE) where exact measurements can be crucial for product quality and safety.

Both models incorporate a large, easy-to-read display, making it convenient for users to view measurement results clearly from a distance. They also feature a wide measurement range, capable of handling a variety of tasks, from basic voltage and current measurements to more complex tasks such as resistance and temperature measurements with appropriate sensors.

The multimeters also stand out with their advanced data logging capabilities. The Fluke 2645A, for instance, can store up to 10,000 readings, allowing for extensive data collection and analysis without the need for constant monitoring. This is beneficial for long-term studies or processes that require continual oversight.

Connectivity is another key characteristic. Both the 2640A and 2645A support USB and RS-232 interfaces, which enable easy integration with computers and other devices for seamless data transfer. This feature is important for engineers who rely on software for data analysis and visualization.

Another notable technology included in these models is the Fluke Connect app compatibility, which allows users to remotely monitor measurements from their smartphones or tablets. This capability provides added flexibility, enabling technicians to work hands-free while still keeping an eye on process measurements.

Additionally, the Fluke 2645A comes with a graphical display option, allowing users to visualize measurement trends over time. This graphical representation can be invaluable for diagnosing issues or observing patterns in data.

In conclusion, the Fluke 2640A and 2645A multimeters offer exceptional performance with their high accuracy, extensive data logging, connectivity options, and advanced visualization capabilities. These features make them indispensable tools for professionals in a variety of fields, ensuring that they can achieve reliable results in their measurement endeavors.