Fluke 2640A user manual Table A-5 A/2645A Trigger Out to Specification

Models: 2645A 2640A

1 353
Download 353 pages 43.78 Kb
Page 216
Image 216

2640A/2645A NetDAQ

Users Manual

Trigger Out

A-8.

Table A-5 provides a summary of the Trigger Out specifications. The Trigger Out output is located on the ALARM/TRIGGER I/O connector, terminals TO and GND.

Table A-5. 2640A/2645A Trigger Out (TO) Specification

Specification

TTL Logical Zero - Trigger Out Set

TTL Logical One - Trigger Out Not Set

Non-TTL Logical Zero - Trigger Out Set

Non-TTL Logical One - Trigger Out Not Set

Pulse Duration (Logic Low)

Isolation

Characteristic

0.8V maximum for an Iout of -1.0 mA (1 LSTTL load)

3.8V minimum for an Iout of 0.05 mA (1 LSTTL load)

1.8V maximum for an Iout of -20 mA

3.25V maximum for an Iout of -50 mA

125µs

None

Master Alarm

A-9.

Table A-6 provides a summary of the Master Alarm specifications. The Master Alarm output is located on the ALARM/TRIGGER I/O connector, terminals MA and GND.

Table A-6. 2640A/2645A Master Alarm (MA) Specification

Specification

TTL Logical Zero - Master Alarm Set

TTL Logical One - Master Alarm Not Set

Non-TTL Logical Zero - Master Alarm Set

Non-TTL Logical One - Master Alarm Not Set

Isolation

Characteristic

0.8V maximum for an Iout of -1.0 mA (1 LSTTL load)

3.8V minimum for an Iout of 0.05 mA (1 LSTTL load)

1.8V maximum for an Iout of -20 mA

3.25V maximum for an Iout of -50 mA

None

A-6

Page 216
Image 216
Fluke 2640A Table A-5 A/2645A Trigger Out to Specification, Table A-6 A/2645A Master Alarm MA Specification