Preparing for Operation 2
Instrument Preparation
External Trigger Wiring for a Group Instrument |
External Trigger Wiring for a group instrument refers to the triggering configuration in which you connect the Master TO (Trigger Out) line to each Slave TI (Trigger In) line and provide a common connection to the GND line for each instrument. This configuration provides improved synchronization of the group instrument when the scanning intervals are 1 second or less.
Figure 2-7 shows a typical wiring connection for a group instrument. Use NetDAQ Logger to configure a group instrument as described in Chapter 3 of this manual.
Controls and Indicators |
The front panel provides a display and a set of control keys; the rear panel provides the power switch and Ethernet status indicators. See Figures