Fluke 2645A, 2640A user manual 20 µV 200 µV 20 mV

Models: 2645A 2640A

1 353
Download 353 pages 43.78 Kb
Page 228
Image 228

2640A/2645A NetDAQ

Users Manual

Table A-25. 2645A DC Voltage Resolution and Repeatability Specifications Resolution

Range

Slow

 

90 mV

3 µV

 

 

300 mV

10 µV

 

 

3V

100 µV

 

 

30V

1 mV

 

 

50V/300V*

10 mV

 

 

* 300V range applies to channels 1 and 11 only.

Fast

6µV

20µV

200µV

2mV

20mV

Range

90mV

300mV

3V

30V

50V

Table A-26. 2645A DC Voltage Accuracy Specifications

 

 

Accuracy, 3σ ± (% input + V)

 

 

 

 

 

 

 

 

18ºC to 28ºC

 

-10ºC to 60ºC

 

 

 

 

 

 

90 Day

1 Year

1 Year

 

 

 

 

 

 

Slow

Fast

Slow

Fast

Slow

Fast

 

 

 

 

 

 

.01%+20 µV

.01%+50 µV

.013%+23 µV

.013%+50 µV

.042%+52 µV

.042%+130 µV

 

 

 

 

 

 

.01%+40 µV

.01%+90 µV

.013%+49 µV

.013%+93 µV

.042%+104 µV

.042%+234 µV

 

 

 

 

 

 

.01%+.3 mV

.01%+.6 mV

.013%+.38 mV

.013%+.64 mV

.042%+.78 mV

.042%+1.56 mV

 

 

 

 

 

 

.01%+4 mV

.02%+8 mV

.013%+4.9 mV

.026%+9.5 mV

.042%+10.6 mV

.084%+20.3 mV

 

 

 

 

 

 

.01%+30 mV

.04%+60 mV

.013%+40 mV

.052%+64 mV

.042%+78 mV

.168%+156 mV

 

 

 

 

 

 

A-18

Page 228
Image 228
Fluke 2645A, 2640A user manual 20 µV 200 µV 20 mV