Specifications

2645A Specifications A

2645A Frequency Measurement Specifications

A-28.

Tables A-37 to A-38 provide 2645A specifications for the frequency measurement function.

Table A-37. 2645A Frequency Accuracy Specifications

 

Frequency Measurement Accuracy, 1 Year, -10ºC to 60ºC

 

 

 

 

Range

 

Resolution

Accuracy + (% input + Hz)

 

 

 

Slow

Fast

Slow

Fast

15 Hz to 900 Hz

900 Hz to 9 kHz

9 kHz to 90 kHz

90 kHz to 900 kHz

1 MHz

0.01Hz

0.1Hz

1 Hz

10 Hz

100 Hz

0.1Hz

1 Hz

10 Hz

100 Hz

1 kHz

0.05%+0.02 Hz

0.05%+0.1 Hz

0.05%+1 Hz

0.05%+10 Hz

0.05%+100 Hz

0.05%+0.2 Hz

0.05%+1 Hz

0.05%+10 Hz

0.05%+100 Hz

0.05%+1 kHz

Table A-38. 2645A Frequency Sensitivity Specifications

Frequency Measurement Sensitivity (Sine Wave)

Frequency Range

Minimum Signal

Maximum Signal

 

 

 

15 Hz to 70 kHz

70 kHz to 100 kHz

100 kHz to 200 kHz

200 kHz to 300 kHz

300 kHz to 1 MHz

100 mV rms

30V rms

 

 

100 mV rms

20V rms

 

 

150 mV rms

10V rms

 

 

150 mV rms

7V rms

 

 

Linearly increasing from 150V

Linearly decreasing from 7V

ms at 300 kHz to 2V rms at 1

rms at 300 kHz to 2V rms at 1

MHz

MHz

 

 