Emerson 5081-P-FF/FI Fisco Csa Intrinsic Safety Installation, 1400288, Hazardous Area

Models: 5081-P-FF/FI

1 28
Download 28 pages 24.89 Kb
Page 22
Image 22
FISCO

22

8

 

7

6

5

4

3

 

 

2

1

 

 

 

 

 

 

 

This document contains information proprietary to

 

 

 

 

 

 

 

 

REVISION

 

 

 

Rosemount Analytical, and is not to be made available

 

 

 

 

 

 

 

 

 

 

 

to those who may compete with Rosemount Analytical.

 

 

 

 

 

LTR

ECO

DESCRIPTION

BY

DATE

CHK

 

 

 

 

 

 

D

D

C

FISCO

CSA INTRINSIC SAFETY INSTALLATION

+PH

THIS DOCUMENT IS

SENSOR

 

CERTIFIED BY

 

CSA

REV

A

 

REV

INFRARED RED

 

REV

REMOTE CONTROL UNIT

 

REV

(RMT PN 23572-00)

 

FOR USE IN

 

REV

 

CLASS I AREA ONLY

REV

REVISIONS NOT PERMITTED

W/O AGENCY APPROVAL

+PH

PREAMP

SENSOR

(NOTE 3)

MODEL 5081-P-FI

XMTR

7

8

9

10

6

 

 

11

5

 

 

12

4

 

 

13

3

 

 

14

2

 

 

15

1

 

 

16

MODEL

5081-P-FI

 

 

XMTR

 

7

8

9

10

6

 

 

11

5

 

 

12

4

 

 

13

3

 

 

14

2

 

 

15

1

 

 

16

HAZARDOUS AREA

IS CLASS I, GRPS A-D CLASS II, GRPS E-G CLASS III

NI CLASS I, DIV 2

GRPS A-D

CLASS II, DIV 2

GRPS E-G

UNCLASSIFIED AREA

CSA APPROVED

UNSPECIFIED

ASSOCIATED APPARATUS

POWER SUPPLY

SUITABLE FOR FISCO

17.5 VDC MAX

SEE NOTE 8 AND TABLE III

 

CSA APPROVED

UNSPECIFIED

ASSOCIATED APPARATUS

POWER SUPPLY

SUITABLE FOR FISCO

17.5 VDC MAX

SEE NOTE 8 AND TABLE III

 

1400288

C

B

CSA APPROVED PREAMP

THAT MEETS REQUIREMENTS

OF NOTE 2

RECOMMENDED CABLE PN 9200273 (UNPREPPED) PN 23646-01 PREPPED

10 COND, 2 SHIELDS 24 AWG SEE NOTE 1

+PH

PREAMP

SENSOR

(NOTE 3)

 

CSA APPROVED PREAMP

 

THAT MEETS REQUIREMENTS

 

OF NOTE 2

 

TB1-

PH

4

5

SENSOR

7

WITH

10

TC

 

MODEL

5081-P-FI

 

 

XMTR

 

7

8

9

10

6

 

 

11

5

 

 

12

4

 

 

13

3

 

 

14

2

 

 

15

1

 

 

16

MODEL

5081-P-FI

 

 

XMTR

 

7

8

9

10

6

 

 

11

5

 

 

12

4

 

 

13

3

 

 

14

2

 

 

15

1

 

 

16

CSA APPROVED

UNSPECIFIED

ASSOCIATED APPARATUS

POWER SUPPLY

SUITABLE FOR FISCO

17.5 VDC MAX

SEE NOTE 8 AND TABLE III

 

CSA APPROVED

 

 

ASSOCIATED APPARATUS

UNSPECIFIED

 

 

 

SUITABLE FOR FISCO

POWER SUPPLY

 

 

 

17.5 VDC MAX

 

 

 

SEE NOTE 8 AND TABLE III

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

UNLESS OTHERWISE SPECIFIED

ITEM

PART NO.

DESCRIPTION

QTY

 

TOLERANCES

 

 

 

 

B

RECOMMENDED CABLE

.XX

-+ .030

ANGLES

+

.XXX

+

.010

 

-

-

 

 

BILL OF MATERIAL

A

4 WIRES SHIELDED

22 AWG. SEE NOTE 1

DIMENSIONS ARE IN INCHES REMOVE BURRS & SHARP EDGES .020 MAX

MACHINED FILLET RADII .020 MAX NOMINAL SURFACE FINISH 125

APPROVALS

DATE

DRAWN B. JOHNSON

5-3-04

Rosemount Analytical,

UnilocUniloc Division

2400 Barranca Pkwy

Irvine, CA 92606

TITLE

A

NOTES ON SHEET 2 OF 2 NOTES: UNLESS OTHERWISE SPECIFICED

WARNING-

SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR

 

 

 

SUITABILITY FOR DIVISION 2.

 

 

 

WARNING-

TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES,

5-6-04

8925

A

 

DISCONNECT POWER BEFORE SERVICING.

 

 

RELEASE DATE

ECO NO.

REV

MATERIAL

FINISH

CHECKED J. FLOCK

5/6/04

PROJECT

J. FLOCK

5/6/04

ENGR APVD

THIS DWG CREATED IN

SOLID EDGE

SCHEMATIC, INSTALLATION

5081-P-FI XMTR CSA

D

DWG NO.

1400288

REV

SIZE

 

A

SCALE

NONE

TYPE

SHEET 1OF 2

8

7

6

5

4

3

2

1

Page 22
Image 22
Emerson 5081-P-FF/FI instruction sheet Fisco Csa Intrinsic Safety Installation, 1400288, Hazardous Area, Unclassified Area