START-UP CONFIGURATION AND CALIBRATION — pH

MODEL 5081-P-FF/FI

START-UP - pH

START-UP CONFIGURATION AND CALIBRATION — pH

To perform a two-point buffer calibration, follow these steps:

1.Place the pH sensor in the first buffer solution.

2.On the remote, press CAL, ENTER, ENTER.

3.Use the arrow buttons to select the correct buffer value. Press ENTER.

4.Rinse the sensor and place it in the second buffer solution. Press ENTER.

5.Use the arrow buttons to select the correct buffer value. Press ENTER.

CALIBRATE

PROGRAM

DIAGNOSE

 

 

CALIbrAtE

CAL bF1

bF 1

bF1 4.01

 

 

 

 

 

Std

 

tEMP AdJ

 

 

 

 

 

Std 7.00

tEMP 25.0

InPut 58.9

SLOPE

OFFSt

GIMP 1000

rIMP 10

5081-P-FF

SoFt

HArd

FAULtS

SLOPE 59.01

nonE

 

CAL bF2

bF 2

dIAGnOStIC CALIBRATEPROGRAM tEMP DIAGNOSEMENU dISPLAY Diag MessageFIGURE 14. Menu Tree for pH bUFFEr Manual backgroundManual backgroundISOPOtntAL Manual backgroundManual background LinE FrEq Manual backgroundManual background PAMP=trAnS Manual backgroundManual background dEFAULt

rOFFSt 060

tAUtO On

tYPE PH

bAUtO On

tCOEF 00.00

LinE 60

FActOrY

bF2 10.01

dIAG OFF

tMAn 25.0

tEMP C

bUFFEr Std

ISO 07.00

 

 

 

 

 

 

 

 

 

 

IMPtC ON

 

 

OUtPUt Cur

 

StAbiLiSE

 

Snr 07.00

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GFH 1500

 

COdE 000

 

tIME 10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GWH 1000

 

 

 

 

deLtA 00.02

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GWL 020

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GFL 010

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL 000

rEF LO

rFH 140

rWH 040

rWL 000

 

 

 

 

 

NO

 

YES

 

 

 

 

 

MENU

Sub-menu

PROMPT

Diag Message

rFL 000

FIGURE 14. Menu Tree for pH

26

Page 26
Image 26
Emerson 5081-P-FF/FI START-UPCONFIGURATION AND CALIBRATION - pH, MENU Sub-menu, Calibrate, Program, Diagnose, Diag Message