Emerson 5081FG instruction manual Accept LOW O2 Slope

Models: 5081FG

1 88
Download 88 pages 49.42 Kb
Page 57
Image 57

Model 5081FG

Instruction Manual

IB-106-5081, Rev. 1.0 May 2005

O2 CONCENTRATION

%

%

LOW CALIBRATION

CHECK GAS O2 VALUE

26020056

Figure 3-29. ACCEPT LOW O2

d.SLOPE

Refer to Figure 3-30. Use this screen to ex- amine the slope calculated from current calibration check. The slope is the amount of cell voltage generated for a given O2 value. After each calibration check, record the slope over the life of the probe. Tracking the slope will indicate if the probe is de- grading. Press NEXT to access the CON- STANT screen.

O2 CONCENTRATION

%

CALCHECK CELL

SLOPE VALUE

26020057

Figure 3-30. SLOPE

e.CONSTANT

Refer to Figure 3-31. Use this screen to ex- amine the cell zero constant calculated from the current calibration check. The constant represents the voltage generated by the cell when no difference exists between the amount of O2 on the reference and process sides of the cell. Note this value for com-

parison against future calibration checks.

Press RESET or EXIT to return to the3 PROCESS DISPLAY screen.

O2 CONCENTRATION

%

CALCHECK CELL

CONSTANT VALUE

26020058

Figure 3-31. CONSTANT