Model 5081FG
Instruction Manual
%
%
26020026
Figure 6-7. Fault 3, Shorted Thermocouple
c.Fault 3, Shorted Thermocouple
The thermocouple connections are shorted. The fault displays as shown in Figure
1.Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approxi- mately 500°C (932°F) may result in this fault.
2.Refer to Figure
3.If the multimeter reading, in voltage mode, is between
4.If the thermocouple is shorted, replace the oxygen probe per paragraph
5.If the thermocouple is not shorted, then replace the faulty analog or CPU board per paragraph
26020027
Figure 6-8. Fault 4, High Probe Temperature
d.Fault 4, High Probe Temperature
The probe’s temperature has exceeded the maximum cell temperature setpoint. The fault displays as shown in Figure
1. If the probe temperature exceeds the | 6 |
maximum cell temperature setpoint, the
2.Verify that the upper cell temperature setpoint is configured as desired under the PROGRAM MENU (see Section 3,
STARTUP AND OPERATION).
Rosemount Analytical Inc. A Division of Emerson Process Management | Troubleshooting |