Emerson 5081FG instruction manual Method 2, For Load Resistance 250 Ohms, Analog Output Device

Models: 5081FG

1 88
Download 88 pages 49.42 Kb
Page 60
Image 60

Instruction Manual

IB-106-5081, Rev. 1.0 May 2005

Model 5081FG

6

7

8

9

10

11

 

5

 

 

 

 

 

12

4

 

 

 

 

 

13

3

 

 

 

 

 

14

2

 

 

4-20 mA

15

1

 

 

16

 

 

 

 

+

 

 

 

 

 

 

MODEL 5081 TRANSMITTER

TERMINAL BLOCK

LOOP CONNECTORS

RL 250

4-20 mA SIGNAL LINE

ANALOG OUTPUT DEVICE

SERIAL PORT & BATTERY LOOP CONNECTORS

USE INTERFACE

00275 0013 ONLY

CHARGER MUST

SERIAL PORT

NOT BE USED IN

 

HAZARDOUS AREAS

 

 

HART

 

COMMUNICATOR

 

LEAD SET

HART COMMUNICATOR

 

REAR PANEL

37420006

Figure 4-1. Signal Line Connections, > 250 Ohms Lead Resistance

a.Method 1, For Load Resistance 250

Ohms

Refer to Figure 4-1 and the following in- struction to connect the HART Communi- cator to a signal line with 250 ohms or more of load resistance.

Explosions can result in death or seri- ous injury. Do not make connections to the HART Communicator's serial port, 4-20 mV signal line, or NiCad re- charger jack in an explosive atmosphere.

Using the supplied lead set, connect the HART Communicator in parallel to the Two- Wire In Situ Oxygen Analyzer. Use any wir- ing termination points in the analog output 4-20 mA signal line.

b.Method 2, For Load Resistance < 250 Ohms

Refer to Figure 4-2and the following steps to connect the HART Communicator to a signal line with less than 250 ohms load resistance.

Explosions can result in death or seri- ous injury. Do not make connections to the HART Communicator's serial port, 4-20 mA signal line, or NiCad re- charger jack in an explosive atmosphere.

1.At a convenient point, break the analog output 4-20 mA signal line and install the optional 250 ohm load resistor.

2.Plug the load resistor into the loop connectors (located on the rear panel of the HART Communicator).

4-2 HART/AMS

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 60