Reference Manual

00809-0100-4697, Rev EA
October 2011
3-15
Rosemount 848T
NOTE
If status is set up to flag failure/warning you will see associated sensor
degraded or failure alert.
Transducer Blocks The transducer block allows the user to view and manage the channel
information. Ther e is one T ra nsducer Block for the eight sensor s that contains
specific temperature measurement data, including:
•Sensor Type
Engineering Units
Damping
Temperature Compensation
• Diagnostics
Maintenance Sensor 8 Degraded Confirm the operating range of Sensor 8
and/or verify the sensor connection and
device environment.
Failed Sensor 1 Failure Verify the Sensor 1 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 2 Failure Verify the Sensor 2 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 3 Failure Verify the Sensor 3 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 4 Failure Verify the Sensor 4 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 5 Failure Verify the Sensor 5 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 6 Failure Verify the Sensor 6 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 7 Failure Verify the Sensor 7 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Sensor 8 Failure Verify the Sensor 8 Instrument process is
within the Sensor range and/or confirm
sensor configuration and wiring.
Failed Body Temperature
Failure Verify that the body temperature is within
the operating limits of this device.
Failed Hardware/Software
Incompatible Contact Service Center to verify the
Device Information
(RESOURCE.HARDWARE_REV, AND
RESOURCE.RB_SFTWR_REV_ALL).
Failed Memory Error Restart the device. If the problem persists,
replace the device.
Failed Electronics Failure Restart the device. If the problem persists,
replace the device.
Alarm Type Active Event Recommended Action