Emerson CMF200A Orienting the Sensor, Recommended sensor orientations, Flow direction arrow

Models: CMF200A

1 26
Download 26 pages 54.18 Kb
Page 12
Image 12

Orienting the Sensor

Step 2 Orienting the Sensor

The sensor will function properly in any orientation if the sensor tubes remain filled with process fluid. Micro Motion recommends orienting ELITE sensors as shown in Table 3.

Table 3 Recommended sensor orientations

Sensor model

Application

Recommended orientation

CMF010

All applications

Tubes flat

 

 

Horizontal pipeline

CMF025, CMF050,

Liquids

Tubes down

and CMF100

 

Horizontal pipeline

 

Gases and slurries

Tubes up

 

 

Horizontal pipeline

CMF200, CMF300,

Liquids

Tubes down

CMF400, and

 

Horizontal pipeline

CMFHC3

 

 

 

Gases

Tubes up

 

 

Horizontal pipeline

 

Slurries

Flag mount

 

 

Vertical pipeline

Flow direction arrow

The sensor has a flow direction arrow (see Figures 1–4) to help you configure the transmitter for flow direction. If possible, install the sensor so that the flow direction arrow matches actual process flow.

Vertical pipeline

If the sensor is installed in a vertical pipeline, liquids and slurries should flow upward through the sensor. Gases may flow upward or downward.

10

Micro Motion® ELITE® Sensors

Page 12
Image 12
Emerson CMF200A Orienting the Sensor, Recommended sensor orientations, Flow direction arrow, Vertical pipeline