Mounting the Sensor

Figure 10 Mounting the temperature extender onto the sensor

Clamping ring screw — when extender is engaged, tighten this to 13–18 in-lbs (1.5–2 N-m)

Feedthrough — note the notches for correct engagement of the extender

Model 2400S transmitter or enhanced core processor — preinstalled on the extended mount

Extended mount — carefully align the notches at the bottom of the extended mount with the notches on the feedthrough to engage properly

Plastic plug — remove and discard

Mounting the electronics of high-temperature and extreme high-temperature sensors

High-temperature and extreme high-temperature ELITE sensors come with a 32(812 mm) pre-installed flexible conduit. This conduit is required for agency approval.

A factory-supplied Model 2400S transmitter, Model 1700/2700 transmitter, core processor, or junction box is connected to the end of the flexible conduit. Mount the electronics to a wall or instrument pole using the supplied bracket (Figure 11).