1-5 SPECIFICATIONS

a.General

Model NGA2000 TO2

Measurement Species

Trace Oxygen

Ranges

0 to 100 ppm (output scalable down to 0-2 ppm fullscale)

Accuracy

±3% of reading or ±0.02% of range (except for ranges 100 ppm:

 

±3% of reading or ±0.05% of range)

Sensitivity

<10 ppb Oxygen

Noise

1% of fullscale, peak to peak

Linearity

±1% of fullscale

Response Time

Typically 90% in less than 20 seconds

Zero Drift

≤±1% of fullscale/24 hours at constant temperature

Span Drift

≤±1% of fullscale/24 hours at constant temperature

Effect of Temperature

0.32% of reading per °F from 70°F

 

(0.58% of reading per °C from 21°C)

Effect of Flow

2% of reading for a flow change of ±250 cc/min (0.5 SCFH)

Operating Temperature

32°F to 113°F (0°C to 45°C)

Power Requirements

+24 VDC ±5%, 10 W max.

 

Ripple and Noise: <100 mV peak to peak

 

Line and Load Regulations: <±1%

b.Sample

 

Sample

Non-flammable (below 100% of the LEL)

 

Flow Rate

0.5 to 1.5 L/min.

 

Supply Pressure

1027 to 1082 hPa - absolute (0.2 to 1.0 psig)

 

Temperature

32°F to 113°F (0°C to 45°C)

 

Particulates

filtered to <0.1 mg/L; non-condensing at ambient temperature

 

Sample Humidity

non-condensing at ambient temperatures

c.

Physical

 

 

Materials in contact with sample ...

Stainless steel, Teflon, Delrin, neoprene

 

Dimensions

See Figure 2-2.Trace Oxygen Analyzer Outline and Mounting

 

 

Dimensions on page 2-3

 

Weight

6.8 kg (15 lbs.)

 

Mounting

Horizontal, external to Platform or custom installed in a panel

 

Case Classification

General Purpose for installation in weather protected area

 

Max. Separation from Platform

1600 m (1 mile)

d.

Gas Connections

 

 

Sample In

1/4 inch O.D. tube fitting

 

Sample Out

1/4 inch O.D. tube fitting

See the Preface Section of the Platform manual for specifications regarding Platform related components.