Instruction Manual

748374-F March 2003

Model NGA2000 TO2

ITEM

PATH

NOTES

Range number selec-

Basic

 

tion

 

 

 

Range lower limits

Expert - Analyzer module setup - Range set-

 

tings

 

 

 

Range upper limits

Expert - Analyzer module setup - Range set-

 

tings

 

 

 

Record service codes

Technical - Service menus – Service history -

 

Analyzer module history - List notes

 

 

 

Reset system

Technical - System setup – System reset

 

Resolution of main

Technical - System setup – Main display con-

 

reading

figuration - Display resolution

 

Sensor current,

Technical - Diagnostics menus - Analyzer mod-

 

temperature

ule diagnostics - Primary variable parameters

 

Sleep mode status

Basic

 

Sleep mode timer

Technical - Diagnostics menus - Analyzer mod-

 

ule diagnostics - Primary variable parameters

 

 

 

Software diagnostics

Technical - Diagnostic menus - Control module

 

diagnostics

 

 

 

Software revision level

Technical - Service menus - Manufacturing data

 

- Control module data

 

 

 

Span calibration

Technical - Diagnostic menus - Analyzer mod-

Highly recommend not to

ule diagnostics - Calibration parameters - User

be conducted in the field

 

span calibration

 

 

Tag number

Technical - Service menus - Manufacturing data

User editable

- Analyzer module data

 

 

Test relay operation

Technical – Listing of all modules – Analog I/O –

 

Select I/O module - Relay status

 

 

 

View sensor calibration

Technical - Diagnostic menus - Analyzer mod-

 

ule diagnostics - Calibration parameters -

 

data

 

Calibration data display

 

 

 

Zero calibration

Technical - Diagnostic menus - Analyzer mod-

Highly recommend not to

ule diagnostics - Calibration parameters - User

be conducted in the field

 

zero calibration

 

 

8-2

Appendix B. User Interface Help

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 44
Image 44
Emerson instruction manual Model NGA2000 TO2