Instruction Manual

748374-F March 2003

Model NGA2000 TO2

From the Run Mode display, press the MENUS softkey to gain access to the Main Menu.

The Basic controls menu is as follows:

 

 

 

 

Basic Controls

 

 

Measurement range number:

 

 

Range upper limit:

CURRENTRNGHI

Range and functional control:

CONTROL

Sleep Mode:

 

 

 

 

Exit Sleep Mode!

 

 

Initiate Quick Start!

 

 

Quick Start:

 

 

 

 

 

 

 

 

 

 

HOME

 

 

ESCAPE

 

 

INFO

 

 

 

 

 

 

 

Figure 3-3. Basic Controls Menu

The analyzer range settings may be config- ured through the Analyzer Module set up menu, under Expert controls and set up.

Analyzer module set up

Range settings . . .

Displayed parameters . . .

 

Analyzer tag:

TAG

 

 

 

 

 

HOME

 

ESCAPE

 

INFO

 

 

 

 

 

Figure 3-5. Analyzer Module Set-up Menu

c.Help Displays

This menu allows the user to view the cur- rent range’s upper limit, Sleep mode, and quick start status. It also allows quick start initiation or exit from sleep mode.

In the figure above, the words in italics are the names of the network variables whose values are in fact shown on the screen.

The Expert controls menu is as follows:

 

 

 

Expert controls

 

 

Measurement range number:

 

 

Range lower limit:

CURRENTRNGHI

Range upper limit:

 

 

Range and functional control:

CONTROL

 

 

 

 

 

 

HOME

 

ESCAPE

 

 

INFO

 

 

 

 

 

 

Figure 3-4. Expert Controls Menu

This menu shows the current range number and range limits.

The Help structure is intended to be an on- line "tutorial," context-sensitive and topic- interconnected, so that the user can practi- cally operate NGA2000 without the need of an instruction manual.

A typical help menu:

Zero/Span calibration help

This allows manual control of the zero and span. Flow zero gas, and make sure the zero gas value is right;

Press the zero key to make the analyzer zero.

Or select Factors and individually adjust the reading on each range. In this way you can make the analyzer read what you want.

Do the same with span gas. If the zero was not a real zero, the span action will change the zero reading>

Set cal gas values in the Calibration gas list screen.

HOME

 

ESCAPE

 

CALINFO

 

INFO

 

 

 

 

 

 

 

Figure 3-6. Typical Help Menu (shown is

Zero/Span Calibration Help)

This is the help screen for the calibration menus.