37
MODEL TCL SECTION 7
PROGRAMMING THE ANALYZER
7.5.3 Procedure – Configuring the Measurement
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. Choose Measurement.
3. The screen at left appears only if you have a dual input 1056. Choose the
Chlorine sensor.
4. The screen summarizes the present configuration for the chlorine sensor.
There are four items: Measure, Units, Filter, and Resolution
To make a change, move the cursor to the desired line and press ENTER.
A screen will appear in which the present setting can be edited. To store
the setting press ENTER.
a. For Measurement choose Total Chlorine.
b. Leave Filter at the default value unless readings are noisy.
For an explanation of terms, see sections 7.5.2.
6. To return to the main display, press MENU then EXIT.
Program
Outputs
Alarms
Measurement
Temperature
Measurement
Configure?
Sensor1
Sensor 2
Sensor 1
S1 Configure
Measure Total Cl
Units: ppm
Filter: 5 sec
Resolution: 0.001 ppm
Measure: