MODEL TCL

SECTION 1

 

DESCRIPTION AND SPECIFICATIONS

1.2 SPECIFICATIONS — SAMPLE CONDITIONING SYSTEM

GENERAL

Enclosure: Fiberglass reinforced polyester, NEMA 3 (IP53) suitable for marine environments

Dimensions: 14.5 x 13.0 x 8.6 in. (369 x 329 x 218 mm)

Mounting: Wall

Ambient Temperature: 32° - 122°F (0 - 50°C)

Ambient Humidity: 0 - 90% (non-condensing)

Power: 115 Vac, 6.9 W, 50/60 Hz; 230 Vac, 7.0 W, 50/60 Hz

Hazardous Location: The TCL sample conditioning sys- tem has no hazardous location approvals.

Pumps:

EN 809:1998

Weight/Shipping Weight: 14 lb/16 lb (6.5 kg/7.5 kg)

SAMPLE REQUIREMENTS

Inlet Connection: compression fitting, accepts 1/4 in. OD tubing

Drain Connection: 3/4 in. barbed fitting (must drain to open atmosphere)

Inlet Pressure: <100 psig (791 kPa abs)

Flow: at least 0.25 gph (15 mL/min)

Temperature: 32 - 122°F (0 - 50°C)

Total Alkalinity: <300 mg/L as CaCO3. For samples con- taining <50 mg/L alkalinity, consult the factory.

SAMPLE CONDITIONING SYSTEM

Reagent: Potassium iodide in vinegar.

Reagent Usage: 5 gallons lasts approximately 60 days.

Reagent Pump: Fixed speed peristaltic pump, about 0.2 mL/min

Sample Pump: Fixed speed peristaltic pump, about 11 mL/min

1.3 SPECIFICATIONS — MODEL 1056 ANALYZER

Case: Polycarbonate NEMA 4X/CSA 4 (IP65).

Dimensions: 6.10 x 6.10 x 5.15 in. (155 x 155 x 131 mm)

Conduit openings: Accepts PG13.5 or 1/2 in. conduit fittings

Display: Monochromatic back-lit LCD. Main character height 0.6 in (15mm). Display is user-programable

Languages: English, German, Italian, Spanish, French, Portuguese

Ambient temperature and humidity: 32 to 131°F (0 to 55°C); RH 5 to 95% (con-condensing)

Storage temperature: -4 to 140°F (-20°C and 60°C)

Power: Code -01: 115/230 VAC ±15%, 50/60 Hz. 10 W. Code -03: 85 to 265 VAC, 47.5 to 65.0 Hz, 15 W (includes four relays)

Equipment protected by double insulation

Hazardous Location Approvals - Applies to analyzer only.

Class I, Division 2, Groups A, B, C, & D

Class Il, Division 2, Groups E, F, & D

Class Ill T4 Tamb= 50°C

Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indi- cators adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively

RFI/EMI: EN-61326

LVD: EN-61010-1

Outputs: Two 4-20 mA or 0-20 mA isolated outputs. Continuously adjustable. Linear or logarithmic. Maximum load 550 ohms. Output dampening with time constant of 5 sec is user-selectable.

Alarms relays (analyzer option -03 only): Four alarm relays for process measurement(s) or temperature. Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured inde- pendently and each can be programmed with interval timer settings.

Relays: Form C, SPDT, epoxy sealed

Relay Contact ratings:

5 A at 28 VDC or 300 VAC (resistive) 1/8 HP at 120/240 VAC.

Terminal Connections Rating: Power connector (3-leads):18-12 AWG wire size. Current output con- nectors (2-leads):24-16 AWG wire size. Alarm relay terminal blocks: 18-16 AWG wire size

Weight/Shipping Weight: (rounded up to nearest lb or nearest 0.5 kg): 1.5 kg (3 lb)/2.0 kg (4 lb)

3

Page 11
Image 11
Emerson TCL instruction manual General, Sample Requirements, Sample Conditioning System

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.