MODEL TCL

 

SECTION 11

 

 

TROUBLESHOOTING

11.4 TROUBLESHOOTING WHEN A WARNING MESSAGE IS SHOWING

 

 

 

 

Warning message

Explanation

Section

 

 

 

Sensor Need Factory Cal

The sensor board was not calibrated at the factory.

11.4.1

 

 

 

Sensor Negative Reading

The chlorine reading is less than -0.5 ppm.

11.4.2

 

 

 

Sensor RTD Sense Open

RTD sensor line is broken or not connected

11.4.3

 

 

 

Sensor Temperature High

Temperature is greater than 155ºC (311ºF)

11.4.4

 

 

 

Sensor Temperature Low

Temperature is less than -20ºC (-4ºF)

11.4.5

 

 

 

11.4.1 Sensor Need Factory Cal

The sensor board was improperly calibrated at the factory. Call the factory for assistance.

11.4.2 Sensor Negative Reading

The analyzer converts the raw current from the sensor to ppm chlorine by subtracting the zero current from the raw current and multiplying the result by a conversion factor. If the zero current is larger than the raw current, the result will be negative.

1.Check the zero current. It should be less than about 50 nA. If it is greater than 50 nA, repeat the zero step.

2.If the zero current is in the correct range, the negative reading might be the result of the raw current or the

sensitivity being too low. A properly operating sensor should generate about 1000 nA for every 1 ppm of total chlorine. Recalibrate the sensor. If necessary, clean or replace the membrane and check the fill solution.

3.Replace the sensor.

11.4.3 Sensor RTD Sense Open

The analyzer measures temperature using a three-wire RTD. See Figure 11.3. The in and return leads are used to measure the resistance of the RTD. The third lead, called the sense line, is connected to the return lead at the sensor. The sense line allows the analyzer to correct for the resistance of the in and return leads and to compen- sate for changes in wire resistance caused by changes in ambient temperature.

1.Check wiring. See Figure 4.4 or 4.5.

2.Disconnect the sense and return wires and check the resistance between them. It should be less than 1Ω. See Figure 4.4 or 4.5.

3.Even though the sense line is open, the sensor is still usable. Use a wire jumper to connect the sense and return terminals on the sensor terminal strip. The temperature reading will no longer be corrected for the lead resistance, nor will the analyzer be able to compensate for changes in ambient temperature. The error could be several ºC or more.

4.Replace the sensor.

11.4.4Sensor Temperature High or Low The sensor RTD is most likely miswired.

1.Check wiring connections.

2.Disconnect the RTD in and return leads and check the resistance between them. See Figure 4.4 or 4.5. The resistance should be close to the values given in Section 11.5.

3.Replace sensor.

62

Page 70
Image 70
Emerson TCL Troubleshooting When a Warning Message is Showing, Explanation Section, Sensor Need Factory Cal

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.