MODEL TCL

SECTION 11

 

TROUBLESHOOTING

11.3 TROUBLESHOOTING WHEN A FAULT MESSAGE IS SHOWING

Fault message

Explanation

Section

 

 

 

Main Board CPU Error

Main board software is corrupted

11.3.1

 

 

 

Main Board Factory Data

Main board factory eeprom data is corrupted

11.3.1

 

 

 

Main Board User Data

Main board user eeprom data is corrupted

11.3.1

 

 

 

Sensor Hardware Error

Missing or bad hardware component

11.3.2

 

 

 

Sensor Board Unknown

Analyzer does not recognize sensor board

11.3.3

 

 

 

Sensor HW-SW Mismatch

Sensor board hardware and software do not match

11.3.3

 

 

 

Sensor Incompatible

Sensor board software is not supported by main board

11.3.4

software

 

 

 

 

 

Sensor Not Communicating

Sensor board is not communicating with main board

11.3.3

 

 

 

Sensor CPU Error

Sensor board software is corrupted

11.3.5

 

 

 

Sensor RTD Open

Temperature measuring circuit is open

11.3.6

 

 

 

S1 Not Detected

No sensor board is connected to sensor 1 terminal

11.3.7

 

 

 

Sensor Factory Data

Sensor board factory eeprom data is corrupted

11.3.8

 

 

 

Sensor EEPROM Write Error

Bad CPU on the sensor board

11.3.8

 

 

 

Sensor User Data

Sensor board user eeprom data is corrupted

11.3.8

 

 

 

Sensor ADC Error

Bad component on the sensor board

11.3.9

 

 

 

Sensor RTD Out of Range

RTD is improperly wired or has failed

11.3.10

 

 

 

11.3.1 Main Board CPU, Main Board Factory Data, and Main Board User Data Errors.

These error messages mean the main board software is corrupted or the eeprom data on the main board is corrupted.

1.Cycle the power off then on.

2.If cycling the power does not help, call the factory. The main board must be replaced. To do this, the analyzer must be returned to the factory.

3.If cycling the power does not help and the fault message was Main Board User Data, reset the analyzer to factory default. See section 7.8. Re-enter user settings and repeat calibration.

11.3.2 Hardware Error.

Hardware error means there is a missing or bad hardware component on the sensor board. The board must be replaced.

11.3.3Sensor Board Unknown, Sensor Board HW (Hardware) or SW (Software) Mismatch, or Sensor Board Not Communicating.

These error messages mean the main board either does not recognize the sensor board or the sensor board and main board are no longer communicating.

1.Verify that the ribbon cable connecting the main board (on the inside of the front panel) and the sensor board are properly seated. Inspect the connecting cable for obvious tears or breaks.

2.If the ribbon cable is properly seated and appears undamaged, the sensor board should be replaced.

60

Page 68
Image 68
Emerson TCL Troubleshooting When a Fault Message is Showing, Fault message Explanation Section, Hardware Error

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.