MODEL TCL

SECTION 10

 

MAINTENANCE

10.2 TOTAL CHLORINE SENSOR

10.2.1 General

When used in clean water, the total chlorine sensor requires little maintenance. Generally, the sensor needs main- tenance when the response becomes sluggish or noisy or when readings drift follow calibration. Maintenance frequency is best determined by experience. If the sensor is used in potable water, expect to clean the mem- brane every month and replace the membrane and electrolyte solution every three months. Sensors used in dirty water require more frequent maintenance and calibration. However, if experience shows the sensor is holding calibration and not drifting appreciably between calibration intervals, the maintenance interval can be extended.

10.2.2 Cleaning the membrane.

Keep the membrane clean. Clean the membrane with water sprayed from a wash bottle. Use a soft tissue to gently wipe the membrane.

10.2.3 Replacing the membrane.

1.Hold the sensor with the membrane facing up.

2.Unscrew membrane retainer. Remove the membrane assembly and O-ring. See Figure 10-2.

3.Inspect the cathode. If it is tarnished, clean it by gently rubbing in the direction of the existing scratches (do not use a circular motion) with 400-600 grit silicon carbide finishing paper. Rinse the cathode thoroughly with water.

4.Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and mem- brane holder pointing up. Fill the cup with electrolyte solution. Set aside.

5.Put a new O-ring in the groove.

6Place a drop of electrolyte solution on the cathode. Invert the membrane assembly and place it over the cathode stem.

7. Screw the membrane retainer back in place.

8.Hold the sensor with the membrane pointing down. Shake the sensor a few times, as though shaking down a clinical thermometer.

10.2.4 Replacing the membrane and electrolyte solution.

CAUTION

Fill solution may cause irritation.

Avoid contact with skin and eyes.

May be harmful if swallowed.

1.Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 10-2.

2.Hold the sensor over a container with the cathode pointing down.

3.Remove the fill plug and allow the electrolyte solution to drain out.

4.Wrap the plug with two turns of pipe tape and set aside. Remove old tape first.

5.Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and mem- brane holder pointing up. Fill the cup with electrolyte solution. Set aside.

6.Hold the sensor at about a 45-degree angle with the cathode end pointing up. Add electrolyte solution (PN 9210438) through the fill hole until the liquid overflows. Tap the sensor near the threads to release trapped air bubbles. Add more electrolyte solution if necessary.

7.Place the fill plug in the electrolyte port and begin screwing it in. After several threads have engaged, rotate the sensor so that the cathode is pointing up and continue tightening the fill plug. Do not overtighten.

8.Place a new O-ring in the groove around the cathode post. Cover the holes at the base of the cathode stem with several drops of electrolyte solution.

9.Insert a small blunt probe, like a toothpick with the end cut off, through the pressure equalizing port. See Figure 10-2.

50

Page 58
Image 58
Emerson TCL instruction manual Total Chlorine Sensor, General, Cleaning the membrane, Replacing the membrane

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.