MODEL 396P and 396PVP | SECTION 7.0 |
| TROUBLESHOOTING |
7.2 TROUBLESHOOTING WITHOUT ADVANCED DIAGNOSTICS.
Table
TABLE
Problem | Probable Cause |
| Remedy |
| |||
|
|
|
|
Meter reads off scale. (Display | Defective preamplifier |
| Replace preamplifier (for code 02 |
reads overrange). |
|
| sensors). For code 01, replace sensor. |
|
|
|
|
| T.C. element shorted |
| Check T.C. element as instructed |
|
|
| in Section 6.1 and |
|
|
| replace sensor if defective. |
|
|
|
|
| Sensor not in process. Sample |
| Make sure sensor is in process with |
| stream is low or air bubbles are |
| sufficient sample stream (refer to |
| present. |
| Section 2.0 for installation details). |
|
|
|
|
| Open glass electrode |
| Replace sensor. |
|
|
|
|
| Reference element open - no contact |
| Replace sensor. |
|
|
|
|
Display reads between 3 and 6 pH | Electrode cracked |
| Replace sensor. |
regardless of actual pH of solution |
|
|
|
or sample. |
|
|
|
|
|
|
|
Meter or display indication swings | T.C. element shorted |
| Check T.C. element as instructed |
or jumps widely in AUTO T.C. Mode. |
|
| in Section 6.1 and replace |
|
|
| sensor if defective. |
|
|
|
|
Span between buffers extremely | T.C. element open |
| Check T.C. element as instructed |
short in AUTO T.C. Mode. |
|
| in Section 6.1 and replace sensor |
|
|
| if defective. |
|
|
|
|
Sluggish or slow meter indication | Electrode coated |
| Clean sensor as instructed in |
for real changes in pH level. |
|
| Sections 6.2 or Section 6.3.2. |
|
|
| Replace sensor if cracked. |
|
|
|
|
| Electrode defective |
| Replace sensor. |
|
|
|
|
Transmitter cannot be standardized. | Electrode coated or cracked |
| Clean Sensor as instructed in |
|
|
| Sections 6.2 or Section 6.3.2 |
|
|
| Replace sensor if cracked. |
|
|
|
|
| Defective preamplifier |
| Replace preamplifier. |
|
|
|
|
Transmitter short spans between | Aged glass electrode or high |
| Replace sensor. |
two different buffer values. | temperature exposure |
|
|
|
|
|
|
| Electrode coated |
| Clean Sensor as instructed in |
|
|
| Section 6.2 or Section 6.3.2. |
|
|
| Replace sensor if cracked. |
| Air bubbles trapped in sensor end |
| Shake the sensor in solution. See |
| between glass bulb and sensor body |
| Section 2.0 for mounting guidelines. |
|
|
|
|
28