SECTION 4.0

MODEL 396P and 396PVP

SECTION 4.0

 

WIRING - MODEL 396PVP

SECTION 4.0

WIRING — MODELS 396P-02 AND 396PVP

Figures 4-1 thru 4-22 provide the guidelines for wiring the sensor to various Analyzer/Transmitter instruments.

To determine which wiring guideline to use, locate the model number of the sensor to be installed.

1.If the cable needs to be extended, use a high qual- ity eleven conductor double shielded instrument cable available from Rosemount Analytical. Refer to Table 4-1 to refer to the appropriate junction box to use and the corresponding wiring details.

NOTE

If the cable is too long, loop up the excess cable. If the cable has to be shortened, cut and terminate each conductor neatly and make sure that the overall (outermost) drain wire is not shorted out with either of the two inner drain wires (shields).

2.Signal cable should be run in a dedicated conduit (preferably an earth grounded metallic conduit) and should be kept away from AC power lines. For your convenience, a wire nut kit is furnished (in a plastic bag wrapped around the cable).

NOTE

For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the extension cable to the instrument must be terminated at earth ground or by using an appropriate metal cable gland fitting that provides a secure connection to the instrument cable.

TABLE 4-1. Wiring Matrix Guide

 

Preamplifier Location

 

 

Analyzer/Transmitter

 

 

 

Sensor

Integral to

Integral to

Remote

Interconnect

1054A,

54, 81, 1055 1181

1050

1003

SCL or

Refer to

Model

Sensor

Analyzer

Junction

(unprepped

1054B, 2054

3081,

1060

1023

SoluCube

Figure

Options

 

 

Box (PN)

PN)

2081

4081, 5081

 

 

 

#

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-55

 



 

 

 



 

 

 

4-3, 4-6,

396PVP-55

 

 

 

 

 

 

 

 

 

4-9

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-50

 



 

 

 



 

 

 

4-4

396PVP-50

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-50

 

 

23309-03

9200254

 







 

4-22

396PVP-50

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-50

 

 

23309-03

9200273 or

 







 

4-8

396PVP-50

 

 

 

23646-01

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-54

 

 

23309-04

9200254



 

 

 

 

4-23

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-54

 

 

23309-04

9200273 or



 

 

 

 

4-20

396PVP-54

 

 

 

23646-01

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-54

 



 

 



 

 

 

 

4-5, 4-18

396PVP-54

 

 

 

 

 

 

 

 

 

4-19

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-55

 

 

23555-00

9200273 or

 



 

 

 

4-7, 4-10

396PVP-55

 

 

 

23646-01

 

 

 

 

 

4-12

 

 

 

 

 

 

 

 

 

 

 

396P-02-()-54

 



 

661-898695

 

 

 

 



4-14

396PVP-54

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

396P-02-54

 

 

23309-04

9200273 or



 

 

 

 

4-11

396PVP-54

 

 

 

23646-01

 

 

 

 

 

4-21

 

 

 

 

 

 

 

 

 

 

 

396P-02-55

 



 

 

 



 

 

 

4-15

396PVP-55

 

 

 

 

 

 

 

 

 

4-16

 

 

 

 

 

 

 

 

 

 

 

396P-02-55

 

 



9200273 or

 



 

 

 

4-17

396PVP-55

 

 

 

23646-01

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16

Page 20
Image 20
Emerson Process Management instruction manual SECTION WIRING - MODELS 396P-02AND 396PVP, 1.Wiring Matrix Guide