6.3 MODEL 396P and 396PVP ORP

MODEL 396P and 396PVP

SECTION 6.0

 

MAINTENANCE

6.3 MODEL 396P and 396PVP ORP

4.3.1Platinum Electrode Check. The platinum elec- trode may be checked as follows: There are two types of standard solutions which may be used to check the ORP electrode/transmitter system.

Type 1: One type of commonly used ORP standard solution is the saturated quinhydrone solution. Refer to Section 5.2.

CAUTION

The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. If contact with skin of clothing is made, immediately rinse with plenty of clean water.

Type 2: A second ORP standard solution is the Ferric- Ferrous Ammonium Sulfate Solution (PN R508-16OZ), and it can be ordered as a spare part; otherwise, it can be prepared from the following recipe: Dissolve 39.2 grams of reagent grade ferrous ammonium sulfate, Fe(NH4)2 (SO4)2 • 6H2O and 48.2 grams of reagent grade ferric ammonium sulfate, FeNH4(SO4)2 • 12H2O, in approximately 700 milliliters of water (dis- tilled water is preferred, but tap water is acceptable). Slowly and carefully add 56.2 milliliters of concentrated sulfuric acid. Add sufficient water to bring the total solu- tion volume up to 1000 ml. This standard ORP solution, although not as simple to prepare as the quinhydrone recipe, is much more stable, and will maintain its milli- volt value for approximately one year when stored in glass containers. This solution (ferric/ferrous ammoni- um sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a saturated KCl/AgCl refer- ence electrode and platinum measuring electrode. Some tolerance in mV values is to be expected due to the rather large liquid reference junction potentials which can arise when measuring this strongly acidic and concentrated solution. However, if the measuring electrodes are kept clean and in good operating condi- tion, consistently repeatable calibrations can be carried out using this standard solution.

6.3.2Cleaning Platinum Electrode. The electrode can be restored to normal operation by simply cleaning the platinum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.