Emerson Process Management 396PVP instruction manual Specifications, Percent Linearity

Models: 396P 396PVP

1 36
Download 36 pages 25.21 Kb
Page 6
Image 6
1.2 SPECIFICATIONS

MODEL 396P and 396PVP

SECTION 1.0

 

DESCRIPTIONS AND SPECIFICATIONS

FIGURE 1-1. Cross Section Diagram of the TUpH Patented Reference Technology

All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH glass bulb. This patented sensor technology ensures superior performance while only requiring minimal maintenance.

1.2 SPECIFICATIONS

Measurements and Ranges

*pH: 0-14

 

ORP: -1500 to 1500 mv

Available pH ACCUGLASS Types

GPLR hemi or flat glass

Wetted Materials

Titanium, Polypropylene, EPDM, glass; platinum (ORP only)

Process Connection

1 in. MNPT front and rear facing threads

Temperature Range

0-100°C (32-212°F)

Pressure Range-Hemi bulb

100-1135 kPa [abs] (0-150 psig)

Pressure Range-Flat bulb

100-790 kPa [abs] (0-100 psig)

Minimum Conductivity

75 μS/cm, nominal; 100 μS/cm

Integral Cable

396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial

 

396PVP: none - must use mating VP cable

Preamplifier Options

396P: Remote or Integral

 

396PVP: Remote only

Weight/Shipping Weight

0.45 kg/0.9 kg (1 lb/2 lb)

 

 

2

PERCENT LINEARITY*

 

pH range

GPLR hemi bulb

 

GPLR flat bulb

 

 

 

 

 

0

- 2 pH

94%

 

 

 

2 - 12 pH

99%

 

98%

 

 

12

- 13 pH

97%

 

95%

 

 

13

- 14 pH

92%

 

 

 

 

 

 

 

 

 

Page 6