3.6 Cable extension
The transducer is connected using ALPHA
Fitting instructions for extending the MSL603 transducer cable are in section 14.
Note 1:
The standard 10 metres of cable should be sufficient for most installations. The maximum recommended additional extension is 20m. Use only cable conforming to the specification given above and ensure any connections are made using high grade screened connectors. For suitable specification connectors refer to Mobrey Measurement customer support. It is the responsibility of the installer to ensure that any extension to the cable must be suitable for its working environment.
Note 2:
The digital input, if used must be connected to switch relay contacts which are insulated to IEC6010 category III (4kV impulse) with no exposed live parts.