7.8 Logging parameters
Logged data can be downloaded and analysed via the RS232 port and Mobrey Measurement Log View Software (via a PC). For copies of this software, please contact Mobrey Measurement Sales.
Note: A full menu structure is shown in section 11.
7.8.1Log Interval - P590
The MSL600 can log the value (not the units) of the parameter that is shown on the middle of the HMI Display. The MSL600 can store up to 4800 records. The interval between each record being stored is programmed in minutes in P590, the Logging interval. The unit takes a reading every 5 seconds and averages the readings over the programmed logging interval.
The unit automatically stores the maximum recorded value that occurred in the previous
Parameter values:
Default | Min | Max |
5 | 0 | 99 |
7.8.2Fast log - P591
The control unit will automatically override the logging interval and log at a rate of one record per minute when the sludge blanket level is above the level in metres (Feet – Imperial) set in the fast Log parameter. If the parameter is set to zero then fast log is not used.
Parameter values (Metric) |
| Parameter values (Imperial) |
| ||
Default | Min | Max | Default | Min | Max |
0.000 | 0.000 | 7.000 | 0.000 | 0.000 | 23.00 |
7.8.3Data Overwrite - P592
If this parameter is set to "On" then old data will be over written when the memory is full. If it is set to "Off" then old data is not over written and the unit stops logging when the memory is full.
Parameter values:
Default | Min | Max |
On | Off | - |
7.8.4Low Mem Alarm - P593
An alarm can be activated if the memory reaches this level when set as a % of full memory.
Parameter Values:
Default | Min | Max |
0 | 0 | 99 |